Product Specifications |
Battery
of Analyzer |
Battery
Backup Time (continuous duty) |
24
hour |
Low Battery indication |
Yes |
Chemistry of Rechargeable Battery |
Li-ion |
Number of Spare Rechargeable
Batteries supplied with the instrument |
0 |
Battery Voltage |
3.7
Volt |
Power Source |
Rechargeable
Battery |
Battery is BIS Approved |
Yes |
BIS Registration number for Battery
under CRS of Meaty |
R-91007129 |
Display
/ Indicator / Storage / Interface |
Display
Type |
LCD |
Record Memory (Minimum) |
50000 |
Display Size |
2.4
inch |
If yes, Capable to display the
geolocation of test inspection / location |
NA |
GPS |
No |
Data Interface |
USB |
Camera |
No |
Record Entry |
Through
Alphanumeric Keyboard |
Availability of audible beeper for
over range, low battery etc. |
Yes |
If yes, Capable to capture the clear
image of test target |
NA |
Record Data Downloadable to PC |
Yes |
Minimum Record Fields |
Full
Information as per the user's requirement such as Driver Name, Vehicle
number, inspector name, test report with time, date, geo-location, instrument
id etc. |
Printer |
Printer |
With |
Type of Printer |
Portable
Dot Matrix Printer with Wireless/without
Bluetooth Connectivity |
Printer Battery is BIS Approved |
No |
Statutory
Requirements |
If
yes, the equipment meets the requirements of the applicable latest Central
Motor Vehicle Rules and all Govt notifications |
Yes |
Immunity to RF interface, meets
requirement as per EN:5008-/1992(emission) and EN:50082-1:1997 (immunity) or
equivalent |
Yes |
The equipment can be used as an
evidential instrument for Traffic Law Enforcement, the instrument is
handheld, portable and easy to operate |
Yes |
Environmental
/ Certification |
Name
of the Lab and Address |
AMIT
LAB Delhi |
Conformity to Dry Heat test at 50
degree C for 16 hours as per IS :9000 latest |
Yes |
Test Report Number and Date |
ATCC2/775
17-07-2021 |
Conformity to Damp heat at 40 degree
C - 95% RH (cyclic test): two cycles of 16+8 hours as per IS: 9000 latest |
Yes |
Conformity to Cold test at - 5 / -10
/ - 15 / -20 degree C (as applicable) for 4 hours as per IS: 9000 latest |
Yes |
Availability of Test Reports from
Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the
specification |
Yes |
Operating
Conditions |
Maximum
Operating Temperature |
65
degrees Celsius |
Minimum Operating Temperature |
-10
degree Celsius |
Relative Humidity Non-Condensing (%) |
95 |
Physical |
Size
(mm x mm x mm) |
165x55x28 |
Weight |
200
grams |
Technology
/ Performance |
RoHS
Compliant |
Yes |
Sensor Technology |
Electrochemical
- Fuel Cell (alcohol specific) |
Sampling modes |
Both
Active and Passive |
Time required for re-calibration |
12
months |
Sampling Error Detection |
Alarm
on wrong sampling |
Barometric Pressure Range (hpa or
millibar) |
640
millibar |
Drift of the breath analyzer in 6
months period (mg / 100 ml BAC) |
1 |
Analytical Response Time (At '0'
level of Alcohol) |
3
second |
Warm up Time (after putting on) |
3
second |
Easily replaceable mouthpiece with
Sputum Trap |
Yes |
Shelf Life of Fuel Cell |
3
years |
Sample Taking Duration |
3
second |
Sampling Method |
Deep
Lung Exhalation Method |
Measuring Range [mg / 100 ml BAC
(Blood Alcohol Concentration)] |
0
to 600 |
Accuracy |
+/-
5 % or better |
Analytical Response Time for positive
result |
8
second |
The Equipment meet RFI requirement as
per EN:5008-1/1992(emission) and EN 50082-1:1997(immunity) or equivalent
requirements |
Yes |
Recovery Time at (1/2) mg/ltr. |
2
second |
Warranty |
Warranty |
1
year |